Error and warning messages
100
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
If a measurement value or arithmetic value lies above or below the value
defined in the SET LIMITS menu, the warning relay is activated and the
„Fault“ LED lights up. The current measurands are present at the analog
output.
In the following cases will the last measured values be kept („frozen“) at the
analog outputs:
1. at all ERROR messages, when the Error relay is activated
2. in standby mode.