Programming
67
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
5.2.4.3
Water detector
If the water falls below the internal water detector level, the message
WATER-ERROR appears. The pump is then switched to filling mode until
the required level is reached again. The message is also output if the
external water detector responds.
ENTER
Don’t make any changes
1
Switch on internal water detector
0
Switch off internal water detector
The purpose of the external water detector is to monitor the water reserve in
the storage tank. Switch the external water detector off if it is not installed, as
otherwise an error message will be output during operation.
ENTER
Don’t make any changes
1
Switch on external water detector
0
Switch off external water detector
5.2.4.4
Hose heating controller
If dew-point temperatures always lie below the ambient temperature even in
winter, and also if the input temperature is controlled externally, it is possible
to do without the hose heating. In this case it is necessary to switch off the
integrated hose heating controller. Otherwise fault reports will be output and
the unit will cease to output measured values.
ENTER
Don’t make any changes
1
Switch on hose heating controller
0
Switch off hose heating controller
EQUIPM.-CONFIG
WATERSEN_IN ? 1/0
EQUIPM.-CONFIG
WATERSEN_EX ? 1/0
EQUIPM.-CONFIG
HOSE-HEATG. ? 1/0