Installation
44
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
3.6
Compressed air connection
In order to operate HYGROPHIL
®
H 4230 it is necessary to install a
compressed air connection that delivers maximum 17.5 Nl/min at a pressure
of 2...5 bar. The pressure required depends on the negative pressure in the
system from which the measurement gas is extracted.
A compressed air conditioner is supplied with HYGROPHIL
®
H 4230 as an
accessory. It contains a reducing valve that can be used to set the
necessary pressure.
The following two diagrams illustrate the interdependencies between
negative pressure and admission pressure as well as between admission
pressure and air consumption.
pressure
pressure