PROFIBUS
106
Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
8.3.4
Output channel
DaAdr (hex)
Identifier
Memory type
Read-out
Format
$31
OutputControlStatus RAM read
hex
Char[3]
$36
Setpoint@
RAM read
dec
Float[3]
$39
OutputChConfig
EEPROM read/write
hex
Record[3]
$3B
Fullscale
EEPROM read/write
dec
Float[3]
$3C
Zeropoint
EEPROM read/write
dec
Float[3]
DaAdr $31: OutputControlStatus
Format: Array [1...3] of Char
7
6
5
4
3
2
1
0
Value OK SignalHigh
SignalLow
Free
Free
Free
Free
Free
Value OK:
1 = Value at analogue output within specified limits
0 = Value at analogue output outside valid range
SignalHigh:
1 = AnalogOut > 100%
0 = AnalogOut
100 %
SignalLow:
1 = AnalogOut < 0%
0 = AnalogOut
0 %
The messages
SignalHigh...LevelLow
can be activated and deactivated
using the corresponding enable bits (DaAdr $39).
DaAdr $36: Setpoint@
Format: Array [1...2] of Float
The value assigned to the analogue output (dimension according to table
$10 or $20).