Operating Instructions HYGROPHIL® H 4230-10 Serie A BA 030520
DaAdr $48: T2 correct config
Format: Char
Before you can write this variable you must first change to standby mode.
0x00 = No correction of HT based on the T2 measurement
0x80 = HT is corrected based on T2 (internal temperature) (default)
8.3.6
2-channel parameters
DaAdr (hex)
Identifier
Memory type
Read-out
Format
$50
PropValveControlStatus
RAM read/write
hex
Record
$54
PumpControlStatus
RAM read/write
hex
Char
$55
PumpConfig
EEPROM read/write
hex
Char
$57
EquipmentControlStatus
RAM read/write
hex
Char
$58
EquipmentConfig
EEPROM read/write
hex
Char
$53
Limit Task Config.
EEPROM read/write
hex
Char [2]
$5a
Limit LOW
EEPROM read/write
dec
Float
$5b
Limit HIGH
EEPROM read/write
dec
Float
DaAdr $50: PropValveControlStatus
Format: Record
Statuschar: Char;
ControlStatusChar: Char;
ON time;
Integer;
OFF time;
Integer;
End;
StatusChar:
7
6
5
4
3
2
1
0
ON
OpenError CloseError Free
Free
Free
Free
Free
ON:
1 = Controller is switched on and working
0 = Controller off
OpenError:
1 = Flow rate too low, valve cannot be opened any further
0 = Valve in working range < maximum
CloseError
1 = Flow rate too high, valve cannot be closed any further
0 = Valve in working range > minimum