R9001440/9
GRAPHICS
TCR Adjustment procedure
Date:10/02/2000
6300
21
TCR Adjustment procedure
At the end of step 1 the program will switch the projector to a full white image and read
the color meter output (if connected) to check the connections. Repeat this check or
stop and restart the program if an error occurs or a bad input parameter has been
given.
If 0 is selected then the TCR Program can be used with any color meter, but only for
step 4 (adjustment of the color temperatures) and step 5 (checking the color tracking).
The type of color meter: LMT, Thoma or Minolta.
Type in type of used color meter and press
ENTER
The TCR Program also requires a perfect adjustment of the projector white panel settings according to
the Panel Adjustment procedure. You can leave the program now if this was forgotten. In the other case
the projector white panel settings will be saved to the factory settings and may no longer be modified
during the rest of the program.
TCR procedure V1.30
Step1: Input projector data
Checking projector and colormeter
Data stream.................................................................
Result..........................................................................
Retry(R), Continue(C) or Abandon(A): _
If no failor has been detected during checking the projector and the colormeter, type
in C (continue) and press
ENTER
to continu the TCR adjustment procedure.
TCR procedure V1.30
Step1: Input projector data
Check for too high gains (y/n)
To preserve the panels for oversaturation, the gain settings will be set below 95%.
Press y(es) to proceed to gain setting.
TCR procedure V1.30
Step1: Input projector data
Check for too high gains (y/n)
Color 1: Gain Driving (<95%): 94.4940 (R)
Color 2: Gain Driving (<95%): 91.2056 (G)
Color 3: Gain Driving (<95%): 91.8306 (B)
Adjustment succeeded..press return
TCR procedure V1.30
Step1: Input projector data
Color meter used:
1: LMT
2: THOMA TMF
3: Minolta CS1000
Your choice:_