9.2.3
Downgrade to an Earlier Version
It is possible to downgrade to an earlier version of the software:
1) Within the server address (see Fig.9.2) change the text “latest” to the desired software
package, e.g., “SW22.”
2) Press
Check for updates
.
3) Follow the procedure described in section 9.2.1 or section 9.2.2.
Note:
Only the software package SW24 or newer can be downgraded to another package.
Software packages down to SW21 are available for update through the Internet.
9.3
Troubleshooting
9.3.1
2250 Light Measurements
If your 2250 Light measurement seems to be wrong, then:
•
Check the cabling, if any
•
Check that the microphone, including preamplifier, is correctly mounted in the top socket
(or correctly connected to the extension cable)
•
Check that the transducer you have mounted on your 2250 Light is selected as the
Transducer Used
parameter, this is also found in the
Setup
menu
•
Check whether the parameters for the selected transducer (
Transducer Used
) are set
correctly, especially the
Microphone Type
parameter. This is found by tapping the Main
Menu icon
, then
Transducers
•
Check that the
Sound Field
and
Windscreen Correction
parameters have been set correctly.
These are found by tapping the Main Menu icon
, then
Setup,
followed by
Input
•
Check if the calibration is OK (make a new calibration using an external calibrator)
9.3.2
SD/CF Cards and USB Sticks
The Logging and Signal Recording software requires that measurements are saved on memory
cards – SD or (with hardware versions 1 – 3) CF. A USB memory stick can be used from
hardware version 4.
Device Format and Speed
In order to ensure reliable operation with the memory device, the format of the file system and
the speed of the device are tested. Upon insertion, the analyzer verifies the format of the
memory device’s file system. If the file system is not optimal for usage with the analyzer, then
you are prompted to format it. You will be guided through the process.
When the format of the file system has been accepted, then the speed of the device is tested.
The test writes data to the device and measures the writing speed. The test takes less than a
minute. If the speed test is successful, then you can proceed using the device, otherwise you
should use another memory device or try reformatting the device (if not done already).
Summary of Contents for 2250 Light
Page 2: ......
Page 10: ......
Page 48: ...38 2250 Light User Manual...
Page 58: ...48 2250 Light User Manual...
Page 116: ...106 2250 Light User Manual...
Page 126: ...116 2250 Light User Manual...
Page 136: ...126 2250 Light User Manual...
Page 168: ...2250 Light User Manual 158...
Page 205: ......
Page 206: ......
Page 207: ......