APPENDIX A
Setup Parameters
145
A.10 Signal Recording
Table A.11
Signal Recording parameters
a
Parameter
Values
Comment
Recording Control
Off
Automatic
Manual Event
Exclude Event
b
External Event
All Events
b
Determines how recording of the measured signal is
controlled
Set to
Automatic
to start the recording when the measurement
is started and record throughout the measurement, only
limited by the
Maximum Duration
Set to
Manual Event
to start recording manually while
measuring when the
Manual Event
pushbutton is pressed,
and record until pressing the pushbutton again, however, take
Maximum
and
Minimum
Duration
into account
Set to
Exclude Event
to start recording manually while
measuring when the
Back-erase
pushbutton is pressed, and
record until pressing the pushbutton again, however, take
Maximum
and
Minimum Duration
into account
Set to
External Event
to start recording using external
equipment connected to the Trigger Input Socket
Set to
All Events
to record while any of the events above are
active, however, take
Maximum
and
Minimum Duration
into
account
If you do not want to record the input signal, then set
Recording Control
to
Off
, to economise on power
Recording Quality
Low
Fair
Medium
High
This setup determines the quality of the recording by adjusting
the sampling rate.
The amount of space required for the recording will depend on
the selected quality and resolution:
Quality Sampling Upper
Memory
Memory
freq.
freq.
16-bit
24-bit
Low
8 kHz
3 kHz
16 KB/s
24 KB/s
Fair
16 kHz
6 kHz
32 KB/s
48 KB/s
Medium
24 kHz 10 kHz
48 KB/s
72 KB/s
High
48 kHz
20 kHz
96 KB/s
144 KB/s
Recorded Signal
b
Input X-weighted
Input C-weighted
Input Z-weighted
Use this parameter to select the frequency weighting of the
recorded signal
Note 1:
The frequency weighting of the recorded signal can
be selected independently of the frequency weighting of the
measurement and the signal at the earphone socket
Note 2:
X
= frequency weighting A or B. ‘A’ requires that the
Broadband (excl. Peak)
parameter is set to
AC or AZ. ‘
B’
requires that the
Broadband (excl. Peak)
parameter is set to
BC or BZ
Summary of Contents for 2250 Light
Page 2: ......
Page 10: ......
Page 48: ...38 2250 Light User Manual...
Page 58: ...48 2250 Light User Manual...
Page 116: ...106 2250 Light User Manual...
Page 126: ...116 2250 Light User Manual...
Page 136: ...126 2250 Light User Manual...
Page 168: ...2250 Light User Manual 158...
Page 205: ......
Page 206: ......
Page 207: ......