CHAPTER 13
Tone Assessment Option BZ-7231 – 1/3-octave Method
121
Fig.13.4
Pop-up window for tone
measurement setup
check
The pop-up window can be de-activated for the rest of the measurement session. To activate it
again, you will have to reload the template or restart your analyzer. This automatic check is
deactivated when the
Tone Assessment
parameter is set to
Off
in the
Setup
menu.
To reset all the relevant parameters to the default setup, tap the
OK
button. (The default setup
includes preset parameters that comply with the selected standard. These are described in Table
13.1.) To measure with the settings you set up manually, tap the
Cancel
button.
Note 1:
Even if you have set the
Tone Assessment
parameter to
On
(
Setup
menu), the
calculation is not carried out until the
Tone
parameter is selected above the spectrum.
13.3.4
Signal Recording
If a valid license for the Signal Recording Option BZ-7226 is present, it is possible to record
the sound while measuring. (Refer to Chapter 12 for information about the Signal Recording
Option.)
Table 13.1
Default parameters used
by the automatic
measurement setup
check
Setup Parameter
Default Value
Standard: ISO 1996-2, Annex D
Adjustment
3 dB
Last band in low frequency range
125 Hz
Last band in middle frequency range
400 Hz
Level difference low
15 dB
Level difference middle
8 dB
Level difference high
5 dB
Summary of Contents for 2250 Light
Page 2: ......
Page 10: ......
Page 48: ...38 2250 Light User Manual...
Page 58: ...48 2250 Light User Manual...
Page 116: ...106 2250 Light User Manual...
Page 126: ...116 2250 Light User Manual...
Page 136: ...126 2250 Light User Manual...
Page 168: ...2250 Light User Manual 158...
Page 205: ......
Page 206: ......
Page 207: ......