CHAPTER 12
Signal Recording (Optional Module)
115
Signal Recordings can be input to the Brüel & Kjær PULSE Multi-Analyzer platform for further
analysis – please contact your local Brüel & Kjær representative for further information.
Note:
When recording signals for further analysis in PULSE, be sure to record the Z-weighted
signal and select
Automatic Gain Control
to
Off
under the
Signal Recording
parameters and
select the
Recording Quality
to match your needs for frequency content – see details on sampling
frequency in Appendix A.
When
Automatic Gain Control
is set to
Off
, then the calibration information is stored in the wave
files – allowing PULSE to analyse the signal recordings, taking the calibration into account.
Note:
Other sound players than BZ-5503 might not be able to amplify “24-bit per sample”
wav-files enough to hear weak sounds. If this is needed, then use 16-bit per sample.