100
2250 Light – User Manual
All markers can be controlled by the stylus in the Profile display.
You can set a
Pre-marker Time
between 0 and 5 seconds. This will start markers 1, 2 or
3 the set number of pre-marker seconds before the point where the
Back-erase
pushbutton
or the
Manual
Event
pushbuttons are pressed.
See section 11.3.2.
10) If you have a license for Signal Recording Option BZ-7226, you can define the settings for
recording the signal while logging under the
Signal Recording
parameters. See Chapter 12
for more information.
11) If you have a license for Tone Assessment Option BZ-7231, you can define its settings
under the
Tone Assessment
parameters. See Chapter 13 for more information.
11.2
Controlling the Measurement
The measurement is controlled in the same way you would control a normal sound level meter
measurement, using
Start/Pause
,
Continue
,
Reset
and
Save
pushbuttons, see Chapter 3 for
more details.
If you have set the
Charge Injection Calibration
parameter to
On
, then CIC is performed at the
start of the measurement and at the end of the measurement. An Exclude marker is set on the
logging profile at the point where the CIC is being performed. The update of the
Total
parameters is stopped while the CIC is being performed. You can see the results of the CIC
under
Total
parameters,
CIC Result 1
and
CIC Result 2
.
Recording the Signal
If you have a license for Signal Recording Option BZ-7226, you can record the input signal
during the measurement by pressing the
Manual
Event
1
or
Manual Event 2
pushbutton (adds a Manual Event marker) or the
Back-erase
pushbutton
(adds an Exclude
marker), or you can record the signal throughout the entire measurement – depending on the
Signal Recording
setting in the
Setup
. Details in Chapter 12.
11.3
Displaying the Results
The Logging measurement screen (or Profile View) includes three tabs at the bottom:
Profile
,
Spectrum
and
Broadband
View (the Logging SLM includes the tabs:
Profile
,
Broadband
and
XL View
). The tabs allow you to choose different ways of displaying the measurement results.
All but the
Profile
tab have been covered previously in Chapters 3 and 10, however a quick
recap is provided below.
The
Broadband
view shows an instantaneous L
AF
readout, with associated bar graph and a
number of measurement parameters. (The first measurement parameter is displayed in a larger
font size for better readability.)
The
XL View
increases the size of the first parameter readout to a 4 digit, full-screen display
(including decimal point).
The
Spectrum
view shows two different spectra parameters being measured simultaneously.
Summary of Contents for 2250 Light
Page 2: ......
Page 10: ......
Page 48: ...38 2250 Light User Manual...
Page 58: ...48 2250 Light User Manual...
Page 116: ...106 2250 Light User Manual...
Page 126: ...116 2250 Light User Manual...
Page 136: ...126 2250 Light User Manual...
Page 168: ...2250 Light User Manual 158...
Page 205: ......
Page 206: ......
Page 207: ......