2250 Light – User Manual
Table 8.2
Content of Notification
Note:
some of the information listed above may not be available, depending on the conditions
when the notification was sent, e.g., no measurement values will be available immediately after
a power on.
8.4.2
Text/SMS Notifications
Notifications via SMS (text message) can be used if you have connected a
GSM/GPRS/EDGE/HSPA modem to the analyzer (see section 8.2.2) and set
Modem
to
Disabled
(a text message cannot be sent if the modem is set to
GPRS/EDGE/HSPA Dialup
).
In addition to sending SMS notifications initiated by alarms, you can send an ‘INFO’ SMS to
the analyzer and ask for an SMS notification, after which you will get a notification within 30
seconds.
You need to setup a
Phone Number
in
Preferences
,
Notifications
,
SMS
and set
Active
to
Yes
for
the SMS. You can also add a Name for the number for easier identification.
You can setup two individual SMS phone numbers for sending the SMS notification to two
recipients.
Set
Notification
to
SMS
and a notification will be sent by SMS to the active phone numbers,
when any of the alarm conditions are met, or if the analyzer receives an ‘INFO’ SMS.
You can test the system by tapping on the Main Menu icon
and selecting “
Send
Notification via SMS
”.
Note 1
: Two to three text messages will be used to send the complete notification – depending
on the content and language settings.
Note 2
: Text messages cannot be sent or received while the analyzer is connected to BZ-5503
or On-line display.
Information Type
Content
Reason for notification
‘
Test
’, ‘
Alarm
’ or ‘
Status
’ followed by a short description
Time of notification
Date
and
Time
Instrument ID
Serial No
. and
Location
Status of current Disk
Free memory
/
Total memory
Status of Internal Battery
Full
,
High
,
Low
,
Critical
,
Charging
or
Unknown
Measurement Status (if available)
Running
,
Paused
or
Stopped
Latest CIC Result (if available)
Passed
,
Background noise too high
,
Ratio outside tolerances
,
Ratio deviates from Ref
.,
Undefined Ref.
, or
Undefined
Current value of LAF (if available)
The value
Summary of Contents for 2250 Light
Page 2: ......
Page 10: ......
Page 48: ...38 2250 Light User Manual...
Page 58: ...48 2250 Light User Manual...
Page 116: ...106 2250 Light User Manual...
Page 126: ...116 2250 Light User Manual...
Page 136: ...126 2250 Light User Manual...
Page 168: ...2250 Light User Manual 158...
Page 205: ......
Page 206: ......
Page 207: ......