38 / 43
51806-001 en/2021-12
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
Error
Possible cause(s)
Measure
Low or no
fl
ow
Blockage within the ultrasonic
bath due to residues from
the contact liquid or textiles
remaining after cleaning.
Thoroughly
fl
ush the piping system inside.
If the desired results are not obtained or
if the pump does not run, the ultrasonic
bath must be sent in for repair.
Please report any malfunctions to us in writing.
6.4
Repair
Repair work may only be carried out by authorised, quali
fi
ed personnel or
by the manufacturer. The manufacturer assumes no liability for unauthorised
interventions on the ultrasonic bath!
6.4.1
Replacing fuses
The ultrasonic bath is protected by 2 electrical microfuses that will activate in the event
of error or overload. If one or both fuses blow, the ultrasonic bath will stop functioning
(the LCD will remain dark).
In general, the ultrasonic bath is defective and must be sent to the manufacturer for repair.
The ultrasonic bath does not need to be opened in order to replace the fuses.
The fuses (2 x microfuses 5 × 20 mm, T4A) are housed in a plug-in module in the mains
socket, which is located on the rear side of the ultrasonic bath.
6.4.2 Determining information on software version and report data
The system data is exclusively used for Service purposes. It must be communicated,
when required, to authorised technical personnel or to the manufacturer.
In the stand-by mode, press and hold the key [
]. A sub-menu will appear with diverse,
current temperature values for the ultrasonic bath (Service menu).
press and hold
T water
17.2 °C
T ch-liquid 25.2 °C
T tank
20.3 °C
water level
o.k.
* S O N O C O O L *
----------------------
12:00:00 17.0°C 50%
12:00:00
17.2°C
12:00:00
With a simple pressing of the key [
] a further sub-menu with system data
(e.g. serial number and software version) may be opened.
T water
17.2 °C
T ch-liquid 25.2 °C
T tank
20.3 °C
water level
o.k.
* S O N O C O O L *
----------------------
SN 3500.00000000.123
R 00.01-Dec 07 2009
ON time 000001:25:11
US time 000001:01:57
CH time 000001:14:20
sw/ar 00266/00000
* S O N O C O O L *
----------------------
With repeated, simple pressing of the key, a third sub-menu [
] will open with
the setting data for the HF generator (FH1)