3.4.2
Time setting
The target runtime can only be edited from stand-by mode. It cannot be edited while
under active operation ( active mode) since the ultrasound continues to be active
while in editing mode.
From the pause mode, only the remaining runtime can be edited.
This setting, however, will not change the runtime speci
fi
cation
(saved value in stand-by mode).
Setting limits: 00:00:10 – 99:59:59 (hh:mm:ss)
Example: taken from stand-by mode
Editing mode ( temperature)
Editing mode ( temperature)
12:00:00 17.0°C 50%
Temperature:
17.2°C
17
12:00:00 17.0°C 50%
Temperature:
17.2°C
17
12:00:00 17.0°C 50%
12:00:00
17.2°C
12:00:00
SET
Editing mode (time)
12:00:00 17.0°C 50%
Time:
12:00:00
12:00:00
3.4.3
Power setting
The power can be changed at any time in editing mode.
Setting values: 25 %, 50 %, 75 % and 100 %.
Example: taken from stand-by mode
Editing mode ( temperature)
Editing mode ( temperature)
12:00:00 17.0°C 50%
Temperature:
17.2°C
17
12:00:00 17.0°C 50%
Temperature:
17.2°C
17
12:00:00 17.0°C 50%
12:00:00
17.2°C
12:00:00
SET
Editing mode (power)
2
12:00:00 17.0°C 50%
Power:
50%
50 %