Horizon Nxt®
Modular Infusion System 8 - 2 6/00
950520 Rev F
Alarm Procedures
An alarm notifies the user that the condition indicated on the Information
Display has occurred and that corrective action must be taken to continue
the infusion process.
When an alarm condition occurs, the following steps are recommended:
1) Press the
H
OLD
key.
2) Press the
M
ENU
key to display on the Information Display the correction
messages contained in the “Correction” column of the Troubleshooting
Guide.
3) Correct the cause of the alarm.
4) Press the
R
UN
key to restart the infusion.
5) Monitor the process according to hospital policy.
There are 3 types of alarms:
1) Operation alarms - Press the
M
ENU
key to see the Help Screen which
contains the guidelines in the “Correction” column of the Alarm
Troubleshooting Guide. Clear the alarm by pressing the
H
OLD
key.
2) System alarm - Press the
H
OLD
key, and press the
R
UN
key to restart the
infusion.
3) Repair Instrument alarm - Press the
H
OLD
key and turn the pump off and
then on again (Similiar to rebooting a personal computer.). If the alarm
recurs, return the pump to a Biomedical Professionial for service.
For assistance, call (800) 627-PUMP.