Horizon Nxt® with MedScan
Modular Infusion System 5 - 3 6/00
950520 Rev F
Hold Extender
The normal Hold Time is 3 minutes. Use this option to extend that
time to a maximum of 120 minutes.
1) Press
H
OLD
key to add time in
3 minute increments, or
2) Press
C
URSOR
keys to select the
hours or minutes and use
D
ATA
keys to change the value
3) Press
E
NTER
key.
"Hold" is displayed on the LEDs. Press the
R
UN
key to cancel the
pause and begin infusing. Press the
E
NTER
key to cancel the
extended pause and return to the standard 3 minute hold. Press the
MENU
key to cancel the pause and access the Menu system.
Micro Rate / Limit
This option is useful in environments where low rates are critical,
such as pediatrics or neonatal nurseries. The maximum rate limit
can be set as low as 5.0 ml/hr or as high as 99.9 ml/hr.
Note
: If the pump is turned on with the Micro Rate Limit activated,
a Teddy Bear shows on the LCD screen for 4 seconds.
1) Press
H
OLD
key
2) Press
M
ENU
key
3) Press
C
URSOR
keys to select Alternate Menu
4) Press
E
NTER
key
5) Press
C
URSOR
keys to select "Micro Rate / Limit"
6) Press
E
NTER
key
Delay = 0 hr 23 min 40 sec
Press HOLD to add 3 min
Cursor to change
Cancel