Horizon Nxt®
Modular Infusion System 7 - 2 6/00
950520 Rev F
“Profile Ready” will appear at top of Information Screen.
Press
R
UN
key to start. “PROFILE RAMP UP” will be displayed on the
Information Screen.
1) Press
E
NTER
key for a graph
2) Press
E
NTER
key for the text
Review/Change PROFILE Data:
Note:
Volume cannot be changed after an infusion is started.
1) Press
M
ENU
key
2) Press
C
URSOR
keys to select “Review/Change PROFILE”
3) Press
E
NTER
key
1) Press
C
URSOR
keys to select Time
2) Press
D
ATA
keys to change the time
3) Press the
C
URSOR
keys to select "VALIDATE PROFILE"
4) Press the
E
NTER
key
5) Press
R
UN
to start
Restart PROFILE:
1) Press
H
OLD
key
2) Press
M
ENU
key
3) Press
C
URSOR
keys to select "Restart PROFILE"
4) Press
E
NTER
key
1) Press
C
URSOR
keys to select "YES"
2) Press
E
NTER
key
3) Press
R
UN
key
Immediate Ramp Down:
1) Press the
M
ENU
key
2) Press the
C
URSOR
keys to select "Immediate Ramp Down"
3) Press the
E
NTER
key
“Immediate Ramp Down” shows on the Information Screen.
Note:
The Immediate Ramp Down time is 30 minutes or less.
RAMP
UP
TEXT
VOLUME DELIV TIME LEFT
0.2 ml 9 H 58 M
The "Restart PROFILE"
option will cancel the
present run. Is this OK?
Yes No
Cursor & Press Enter
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com