Horizon Nxt® with MedScan
Modular Infusion System 5 - 5 6/00
950520 Rev F
Piggyback Callback
Piggyback Callback, if activated, continues the primary infusion at
a KVO rate and causes the pump to alarm. This feature may be
accessed during the Run or Hold states.
1) Press
M
ENU
key
2) Press the
C
URSOR
key to select Alternate Menu
3) Press
E
NTER
key
4) Press
C
URSOR
keys to select “Piggyback Callback”
5) Press
E
NTER
key
1) Press
C
URSOR
keys to select On
2) Press
E
NTER
key
Note:
Piggyback Callback will
return to the
Off
position when the
pump is turned off.
PROFILE Mode
The Profile Mode is described in Chapter 7 (Special Modes).
PROGRAM Mode
The Program Mode is described in Chapter 7 (Special Modes).
q(x)° Schedule
This option allows the programming of periodic antibiotics,
chemotherapy or other IV infusions. For example, antibiotics are
written to be infused "
q
6° " (every 6 hours). It is important to have
a maintenance IV infusing. This option can only be enabled by
Biomedical personnel.
Note:
A cassette must be loaded in the pump to program this
feature. The piggyback option is not available while using
this feature. The
q(x)°
option cannot be used while in DOSE,
PROFILE or PROGRAM Modes.
1) Press
H
OLD
key
2) Press
M
ENU
key
3) Press
C
URSOR
keys to select q(x)° Schedule
4) Press
E
NTER
key
When Piggyback Callback
is ON, KVO will alarm as
soon as the Piggyback is
done.
Set Callback: Off On