Horizon Nxt® with MedScan
Modular Infusion System 5 - 4 6/00
950520 Rev F
1) Press right
C
URSOR
key to select
On
2) Press down
C
URSOR
keys to select
Rate Limit
3) Press
D
ATA
keys to set a rate limit
4) Press
E
NTER
key
Note:
If a rate exists or is entered that is higher than the Rate Limit
set, an audible signal occurs. The user will need to lower the
rate or exit this feature.
1) Press
R
UN
key to start the
infusion.
Monitor Docking
Monitor Docking allows the Horizon Nxt® with MedScan to
communicate with select monitoring systems to display current
infusion parameters. Parameters displayed may vary depending on
which monitoring system is being used. Refer to Monitor Docking
Guide for operations and troubleshooting.
Set-up
Clamp the Serial Data Communication IR (Infrared) Module over
the Serial Data Communication fluid resistant window on the back
of the Horizon Nxt pump whose information you wish displayed.
1) Grasp the Serial Data Communication IR Module with index
finger on top and thumb on grooved surface on the bottom.
2) Hook the top of the unit onto the pump, and with thumb press the
grooved surface to clamp the bottom of the unit on the pump.
Occlusion Warning
Occlusion Warning is designed to alert the clinician of
downstream resistance. An audible warning (three beeps) and
visual a indication will occur 5 times at 5-second intervals before
alarming "Occlusion." This option must be turned on by the
Biomedical Department.
Set Micro Status: Off On
Press ENTER when done
Rate Limit = 59.9 ml/hr
MICRO
Time Left = 7 hr 34 min
Total Infused = 24.1 ml
Occlusion Limit = 300 mmHg