background image

Horizon Nxt® with MedScan
Modular Infusion System                                                              5 - 4                                                                                                 6/00
950520 Rev F

1)   Press right 

C

URSOR

 key to select 

On

2)   Press down 

C

URSOR

 keys to select 

Rate Limit

3)   Press 

D

ATA

 keys to set a rate limit

4)   Press 

E

NTER

 key

Note:

  If a rate exists or is entered that is higher than the Rate Limit 

set, an audible signal occurs.  The user will need to lower the 
rate or exit this feature.

1)   Press 

R

UN

 key to start the 

infusion.

Monitor Docking

Monitor Docking allows the Horizon Nxt® with MedScan to 
communicate with select monitoring systems to display current 
infusion parameters.  Parameters displayed may vary depending on 
which monitoring system is being used.  Refer to Monitor Docking 
Guide for operations and troubleshooting.

Set-up

Clamp the Serial Data Communication IR (Infrared) Module over 
the Serial Data Communication fluid resistant window on the back 
of the Horizon Nxt pump whose information you wish displayed.

1)   Grasp the Serial Data Communication IR Module with index 

finger on top and thumb on grooved surface on the bottom.

2)   Hook the top of the unit onto the pump, and with thumb press the 

grooved surface to clamp the bottom of the unit on the pump.

Occlusion Warning 

      

      Occlusion Warning is designed to alert the clinician of 

downstream resistance.  An audible warning (three beeps) and 
visual a indication will occur 5 times at 5-second intervals before 
alarming "Occlusion."  This option must be turned on by the 
Biomedical Department.

Set Micro Status:      Off On

Press ENTER when done

Rate Limit = 59.9 ml/hr

MICRO

Time Left  =  7 hr 34 min
Total Infused  =  24.1 ml
Occlusion Limit = 300 mmHg

Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Summary of Contents for HORIZON Nxt

Page 1: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Page 2: ...ual may not in whole or in part be copied photocopied reproduced translated or converted to any electronic or machine readable form without prior written consent of B Braun Medical Inc Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 3: ...back Callback 3 3 Changing the Rate 3 3 During Infusion titration 3 3 With Pump on Hold 3 4 Chapter 4 MENUS 4 1 Main Menu 4 1 Alternate Menu 4 1 Change Alarm Volume 4 2 Clear Total Infused 4 2 Dose Mode 4 2 Occlusion Limit 4 3 Alternate Menu Items 4 4 Calculate BSA 4 4 Set LCD Contrast 4 4 Chapter 5 CUSTOM MENU OPTIONS 5 1 Check Battery Level 5 2 Hold Extender 5 3 Micro Rate Limit 5 3 Monitor Dock...

Page 4: ...le Parameters 7 1 Data Entry 7 1 Review Change PROFILE Data 7 2 Restart PROFILE 7 2 Immediate Ramp Down 7 2 Exit PROFILE Mode 7 3 Program Mode 7 4 Data Entry 7 4 Review Change PROGRAM 7 5 Restart PROGRAM 7 6 Skip PROGRAM Periods 7 6 Exit PROGRAM Mode 7 6 Chapter 8 MAINTENANCE 8 1 Cleaning 8 1 Alarm Procedures 8 2 Alarm Troubleshooting Guide 8 3 Cautions 8 5 Service and Performance Information 8 7 ...

Page 5: ...by operation of law or otherwise including but not limited to any implied warranties of merchantability or fitness since handling storage and cleaning of these devices as well as factors relating to other matters beyond B Braun Medical s control directly affect these devices and the results obtained from their use B Braun Medical shall not be liable for any incidental or consequential loss damage ...

Page 6: ...g Because some infusions require multiple flow rates during the delivery cycle the Horizon Nxt is equipped with two automatic multiple delivery rate modes The Profile Delivery mode incorporates a standard ramped delivery regimen and the Programmable Delivery mode can be set up to sequentially deliver up to nine user selected volumes of fluid each at a user selected rate Both modes are designed for...

Page 7: ...s less than 0 1 mL The administration of drugs is restricted to those anesthetic and analgesic drugs labeled for continuous epidural administration Warning Serious injury to the patient may result from epidural administration of drugs other than those specifically labeled for epidural use Blood Infusion The Horizon Nxt can be used for infusion of blood and blood products For blood infusion a Horiz...

Page 8: ...time Program Mode offers maximum flexibility in customizing IV fluid administration Up to 9 separate periods of fluid volume can be infused sequentially q x Mode can be used for the periodic infusion of antibiotics chemotherapy or other IV infusions according to a schedule set by the user Assure that there is a maintenance IV infusing in the intervals between pump infusions Operating States There ...

Page 9: ...se Indicator 16 Piggyback Rate Display 6 Power Key 17 Piggyback Volume Display 7 Primary Indicator Key 18 Information Screen 8 Piggyback Indicator Key 19 Primary Rate Display 9 Primary Run Key 20 Primary Volume Display 10 Piggyback Run Key 21 Door Lever 11 Data Keys 7 18 11 12 13 14 15 19 17 16 20 3 2 1 9 10 21 4 5 6 8 18 19 16 20 17 Artisan Technology Group Quality Instrumentation Guaranteed 888 ...

Page 10: ...ock Out Switch 2 Serial Communication Data Port Figure 3 Inside the Door 1 Cassette Alignment Pins 2 Air In Line Detector 3 Free Flow Protection Clip 1 2 B Braun Medical Carrollton TX 75006 1 2 3 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 11: ...to be plugged into each other with only one pump plugged into a wall outlet On the side of each Horizon Nxt pump is an electrical outlet Only another Horizon or Horizon Nxt pump should be plugged into that outlet Caution To daisy chain more than 5 pumps verify that the total leakage current as a system does not exceed 100 uA Battery Operation The Horizon Nxt has been equipped with an internal rech...

Page 12: ... roller clamp 6 Position the cassette over the Cassette Alignment Pins with the metal disc toward the pump Thread the tubing through the Air In Line Detector Insert and press the Free Flow Protector Clip into the receptacle Warning Unrestricted fluid flow may occur if the Free Flow Protector Clip is not properly installed in the receptacle 7 Close the door of the pump 8 Open all clamps on the pump...

Page 13: ... the pump will alarm and go into the KVO state For Piggyback Fluids The Piggyback option can only be enabled by Biomedical personnel The piggyback option is not available while operating the pump in the Dose Mode Caution Do not attempt to infuse both fluids simultaneously using this method 1 Prepare IV fluids according to the set package Use primary sets with check valves 2 Connect the piggyback s...

Page 14: ...The infusion will now be delivered at the primary rate Make sure the primary tubing is not clamped Note If the secondary container is not empty the remaining secondary fluid will be delivered at the primary flow rate Piggyback Callback This feature causes the pump to alarm and continue the infusion at a KVO rate after the programmed piggyback volume is infused Piggyback Callback must be selected i...

Page 15: ...on System 3 4 6 00 950520 Rev F 1 Press the HOLD key 2 Press the DATA keys to change the rate 3 Press the RUN key to start the infusion Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 16: ...rmation Screen Dose Mode calculates the rate or dose information necessary for use with many critical drug infusions Occlusion Limit sets the occlusion pressure limit to 75 100 200 300 400 or 500 mmHg Alternate Menu lists options used less often by the user Exit from Menu returns to the Information Screen Alternate Menu Calculate BSA calculates the body surface area BSA Set LCD Contrast changes th...

Page 17: ...layed on the Information Screen Note The volume can be cleared while the pump is running 1 Press MENU key 2 Clear Total Infused is already selected 3 Press ENTER key 1 Press ENTER key 2 The total volume displayed on the Information Screen goes to zero The pump will return to the previous Hold or Run state Dose Mode The Dose Mode is fully described in the Dose Mode Chapter Change Alarm Volume Low M...

Page 18: ... beeps and a visual indication in the LED display to let the user know an occlusion is occuring downstream from the pump This warning occurs 5 times at 5 second intervals before alarming Occlusion This feature is designed to alert the clinician of downstream resistance before an alarm occurs which stops the infusion Caution The higher the pressure limit the less sensitive the pump is to changes in...

Page 19: ...CURSOR keys to select Ht and Wt values 2 Press the DATA keys to enter the information 3 The BSA value is displayed 4 Press ENTER to exit Set LCD Contrast This option allows the user to change the readability of the Information Screen 1 Press MENU key 2 Press CURSOR keys to select Set LCD Contrast 3 Press ENTER key 1 Press right and left CURSOR keys to vary the contrast 2 Press ENTER key to validat...

Page 20: ...tion must be turned on by the Biomedical Department Piggyback Callback continues the primary infusion at a KVO rate and alarms once the piggyback is complete PROFILE Mode incorporates a ramped delivery regimen based on the patient s specific needs PROGRAM Mode lets the user program a maximum of nine selected volumes of fluid q x Schedule allows the programming of periodic IV infusions Schedule Nex...

Page 21: ...rnate Menu 3 Press the MENU key 4 Press CURSOR key to select Check Battery Level 5 Press ENTER key One of the three graphs below will be displayed Discharging The pump is unplugged or running on battery Charging The pump is not fully charged Fully Charged The pump is 95 or more charged Press the ENTER key to return to the Hold or Run state Check Battery Level DISCHARGING EXIT Check Battery Level C...

Page 22: ...hold Press the MENU key to cancel the pause and access the Menu system Micro Rate Limit This option is useful in environments where low rates are critical such as pediatrics or neonatal nurseries The maximum rate limit can be set as low as 5 0 ml hr or as high as 99 9 ml hr Note If the pump is turned on with the Micro Rate Limit activated a Teddy Bear shows on the LCD screen for 4 seconds 1 Press ...

Page 23: ...rial Data Communication IR Infrared Module over the Serial Data Communication fluid resistant window on the back of the Horizon Nxt pump whose information you wish displayed 1 Grasp the Serial Data Communication IR Module with index finger on top and thumb on grooved surface on the bottom 2 Hook the top of the unit onto the pump and with thumb press the grooved surface to clamp the bottom of the u...

Page 24: ...e The Program Mode is described in Chapter 7 Special Modes q x Schedule This option allows the programming of periodic antibiotics chemotherapy or other IV infusions For example antibiotics are written to be infused q 6 every 6 hours It is important to have a maintenance IV infusing This option can only be enabled by Biomedical personnel Note A cassette must be loaded in the pump to program this f...

Page 25: ...s the DATA keys to set the infusion rate in ml hr 1 Press CURSOR keys to select Dose Vol 2 Press the DATA keys to set the volume of each dose for example 4 doses each 50 ml Note The Dose Volume of each dose has to be greater than 1 ml but not more than 2000 ml 3 Press the ENTER key To change the time 1 Press up CURSOR key to select Present Time is 2 Press DATA keys to set the correct time Note The...

Page 26: ...ows on the LED s No fluid is infusing during this period 1 Press ENTER key for the data screen 2 Follow the steps above to change the data if necessary 3 Press ENTER key At the q x Start time the pump will begin the infusion During the infusion SCHd run shows on the LED s The number of doses remaining is shown on the LCD After the infusion is complete the pump will automatically return to the wait...

Page 27: ...System 5 8 6 00 950520 Rev F Note The pump must be in the Hold state 1 Press MENU key 2 Press CURSOR keys to select Exit q x Schedule 3 Press ENTER key Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 28: ...ge the Delay Note The time is displayed in military time Use 00 00 to represent midnight and 12 00 to represent noon Either Press CURSOR keys to select hr or min Press DATA keys to set the hours and minutes or Press CURSOR keys to select Start at Press DATA keys to change the time 1 Press CURSOR keys to select Validate 2 Press ENTER key The information is displayed on the Information Screen dLY is...

Page 29: ...es of infusion Caution To prevent inadvertent infusion of drugs not indicated for epidural administration use only Horizon microbore administration sets without injection sites Following an occlusion alarm a small amount of fluid may be delivered Use of a microbore set assures this is less than 0 1 mL The administration of drugs is restricted to those anesthetic and analgesic drugs labeled for con...

Page 30: ...s HOLD key 2 Press MENU key 3 Press CURSOR keys to select Set Time and Volume 4 Press ENTER key 1 Press CURSOR keys to select hr or min 2 Press DATA keys to enter the correct hours and minutes 3 Press CURSOR keys to select ml 4 Press DATA keys to enter the volume 5 Press CURSOR keys to select VALIDATE 6 Press ENTER key 7 Press the RUN key to start the infusion Set Time and Volume Rate 125 00 ml hr...

Page 31: ...D or MIN resolution Note MAX is the default setting MIN displays more data points compressed along the horizontal axis for a representation of a long term trend MAX displays less data points in the same area for a short term trend Resistance Factor RF Note This feature needs to be activated by the Biomedical Technician The Resistance Factor is a numeric value displayed and updated during an infusi...

Page 32: ...0 950520 Rev F This feature returns the pump to the state it was in before pressing the Menu key 1 Press the CURSOR keys to select Exit from menu 2 Press the ENTER key Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 33: ...een or To change to a new medication 1 Press the CURSOR key to select Change 2 Press ENTER key or To Exit DOSE Mode 1 Press CURSOR key to select Exit Mode 2 Press ENTER key Drug Menus 1 Press the CURSOR key to select a category 2 Press ENTER key An alphabetical list of medications is shown 1 Press CURSOR keys to select the medication 2 Press ENTER key To clear the drug name DOPAMINE is the current...

Page 34: ... be done here The concentration can be changed later See Data Screen Data Screen Body Weight 1 Press CURSOR keys to select the Body Weight value if shown 2 Press DATA keys to change the value Calculated Parameter Concentration Dose and Rate are interdependent on each other for their Pressure will be set to 750 mmHg and Micro Rate will be enabled Use sets labeled for Epidural use only DOPAMINE 400 ...

Page 35: ...the calculated parameter Enter Data 1 Press DATA keys to change value 2 Press CURSOR key to select 3 VALIDATE 4 Press ENTER key 5 Press RUN key to start the infusion Rate and Volume data is transferred to the displays The name of the medication and DOSE are shown on the Information Screen Change SETUP Parameters Critical medications require parameters for calculation of doses for example mcg kg mi...

Page 36: ... infuse at the old rate An intermittent alarm will continue to sound to notify the user that a rate change has been attempted and cancelled Pressing any key will silence the alarm Titrate Dose 1 Press ENTER key 2 Press CURSOR keys to select the Dose value 3 Press DATA keys to change the values 4 Press CURSOR keys to select VALIDATE 5 Press ENTER key Monitor the infusion as indicated by hospital po...

Page 37: ... Press Data keys to change the values 3 Press Cursor keys to select VALIDATE Parameters 4 Presss Enter key Note The minimum ramp time is 10 minutes the maximum ramp time 120 minutes The largest value for maximum rate is 400 ml hr Data Entry 1 Press MENU key 2 Press CURSOR keys to select PROFILE Mode 3 Press ENTER key 1 Press DATA keys to enter the Volume 2 Press CURSOR keys to select time 3 Press ...

Page 38: ...OR keys to select VALIDATE PROFILE 4 Press the ENTER key 5 Press RUN to start Restart PROFILE 1 Press HOLD key 2 Press MENU key 3 Press CURSOR keys to select Restart PROFILE 4 Press ENTER key 1 Press CURSOR keys to select YES 2 Press ENTER key 3 Press RUN key Immediate Ramp Down 1 Press the MENU key 2 Press the CURSOR keys to select Immediate Ramp Down 3 Press the ENTER key Immediate Ramp Down sho...

Page 39: ... 1 Press HOLD key 2 Press MENU key 3 Press CURSOR keys to select Exit PROFILE Mode 4 Press ENTER key 1 Press CURSOR keys to select YES 2 Press ENTER key The pump will return to the standard Hold state The Exit PROFILE mode option will cancel the present run Is this OK Yes No Cursor Press Enter Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 40: ...ect PROGRAM Mode 3 Press ENTER key Data Entry 1 Press CURSOR keys to select Rate or Volume 2 Press DATA keys to change the values 3 Press CURSOR keys to select the next period 4 Repeat until all information is entered up to 9 periods Note VALIDATE cannot be selected when 0 0 data is present in the middle of a Program for example NOT ALLOWED PROGRAM Operation was Interrupted Resume current PROGRAM ...

Page 41: ... the rate in that period is much less compared to the other periods Review Change PROGRAM Note A Program cannot be changed after it has been started 1 Press MENU key 2 Press CURSOR keys to select Review Change PROGRAM 3 Press ENTER key 1 Press CURSOR keys to select rate 2 Press DATA keys to change the rate values 3 Press ENTER key Note Only rate can be changed after an infusion has started 4 Press...

Page 42: ... the DATA keys for VTBD to remove the volume 2 The bar on the graph moves to show the volume skipped 3 Press ENTER key 4 Press RUN key Upon completion of the program the pump enters the KVO state Press the HOLD key Exit PROGRAM Mode 1 Press HOLD key 2 Press MENU key 3 Press CURSOR keys to select Exit PROGRAM Mode 4 Press ENTER key The Restart PROGRAM option will cancel the present run Is this OK Y...

Page 43: ...0 5 Sodium Hypochlorite solution one part bleach with nine parts water The solution is most effective when prepared weekly and allowed to remain on the device for approximately 10 minutes prior to wiping off Please be sure NOT to use abrasive cleaning tools or pressurized spraying devices for these may damage the pressure transducer or cause fluid intrusion Warning To avoid mechanical or electroni...

Page 44: ...ss the RUN key to restart the infusion 5 Monitor the process according to hospital policy There are 3 types of alarms 1 Operation alarms Press the MENU key to see the Help Screen which contains the guidelines in the Correction column of the Alarm Troubleshooting Guide Clear the alarm by pressing the HOLD key 2 System alarm Press the HOLD key and press the RUN key to restart the infusion 3 Repair I...

Page 45: ...lter blocked Tubing kinked IV positional or infiltrated Catheter or vein too small Check occlusion limit Hold Time Exceeded HOLd Press Hold key three minute time period is restarted KVO flashing LED s Pump is operating in Keep Vein Open state Infusion is complete Set volume to be delivered Restart or turn off the pump Low Battery PLUG IN Battery too low plug in AC System Error SYS Err Turn the pum...

Page 46: ...m 8 4 6 00 950520 Rev F Set Piggyback Volume Set a Piggyback volume Set Primary Rate Set a Primary rate Set Primary Volume Set a Primary volume Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 47: ...t of fluid may be delivered Use of a microbore set assures this is less than 0 1 mL Serious injury to the patient may result from epidural administration of drugs other than those specifically labeled for epidural use It is not recommended to stack more than five pumps on one I V pole Use only with B Braun Medical Horizon I V Sets Perform testing and or preventive maintenance as defined in the pro...

Page 48: ...ot steam autoclave or immerse the pump in any fluids or cleaning solutions A possible electrical shock hazard exists if the case of the pump is opened or removed Refer all servicing to authorized service personnel Do not use visibly damaged pumps or pumps with bent damaged or missing components Pumps should be tested for proper performance semiannually Federal U S A law restricts this device to sa...

Page 49: ...ion operation training service training and service manuals may be obtained from the manufacturer by contacting B Braun Medical Inc 1601 Wallace Drive Suite 150 Carrollton Texas 75006 USA Attn Manager of Service Or call 800 627 PUMP Product complaints may be sent to the Manager of Quality Assurance at the above address With each complaint please include 1 The pump s serial number 2 Software Revisi...

Page 50: ...5 Plug Hospital Grade 3 pin Pole Size Range 0 75 to 1 25 Battery 12 Volt sealed lead acid battery for ambulatory use Run time is dependent on the state of bat tery charge and the rate of fluid delivery Rate Range Standard Primary and Piggyback 0 1 to 999 9 mL hr Micro Primary and Piggyback 0 1 to 99 9 mL hr Volume to be Infused Range 0 1 to 9999 9 mL Occlusion Pressure 75 100 200 300 400 500 and 7...

Page 51: ... Alarm 0 09 cc Hold Time Exceeded Alarm Activated after three minutes on hold without user interaction deactivated if the door of the pump is open Memory Permanent data retention after the pump is turned off unless the battery is depleted or dis connected or unless the Data Retention De faults have been otherwise selected in the Biomed Options Artisan Technology Group Quality Instrumentation Guara...

Page 52: ...Amiodarone 150 mg in 100 ml 900 mg in 500 ml 0 mg in 0 ml mg min Amrinone 500 mg in 500 ml 500 mg in 250 ml 0 mg in 0 ml mcg kg min Bretylium 2 g in 500 ml 4 g in 500 ml 0 g in 0 ml mg min Cisplatin 0 mg in 1000 ml 0 mg in 500 ml 0 mg in 250 ml mg m2 hr Cyclophosphamide 0 mg in 0 ml 0 mg in 0 ml 0 gm in 0 ml mg kg min Dexamethasone 0 mg in 0 ml 0 mg in 0 ml 0 mg in 0 ml mg kg hr Diltiazem 125 mg i...

Page 53: ...n Etoposide VP 16 0 mg in 0 ml 0 mg in 0 ml 0 mg in 0 ml mg m2 hr Fentan mcg kg min 50 mcg in 1 ml 2 5 mg in 50 ml 2 5 mg in 50 ml mcg kg min Fentan mcg kg hr 50 mcg in 1 ml 2 5 mg in 50 ml 2 5 mg in 50 ml mcg kg hr Fluorouracil 5 FU 0 mg in 0 ml 0 mg in 0 ml 0 mg in 0 ml mg kg hr Furosemide 100 mg in 50 ml 0 mg in 0 ml 0 mg in 0 ml mg min Heparin units hr 20 000 units in 500 ml 25 000 units in 25...

Page 54: ...00 ml 200 mg in 250 ml 0 mg in 0 ml mg min Lidocaine 2 gm in 500 ml 4 gm in 500 ml 0 gm in 0 ml mg min Magnesium Sulfate 5000 mg in 250 ml 5000 mg in 500 ml 0 mg in 0 ml g hr Methotrexate 0 mg in 0 ml 0 mg in 0 ml 0 mg in 0 ml mg kg min Midazolam 25 mg in 50 ml 50 mg in 50 ml 0 mg in 0 ml mg hr Milrinone 10 mg in 10 ml 20 mg in 100 ml 0 mg in 0 ml mcg kg min Morphine 10 mg in 50 ml 10 mg in 1 ml 0...

Page 55: ... in 500 ml 0 units in 0 ml millU min Phenylephrine 30 mg in 500 ml 50 mg in 500 ml 0 mg in 0 ml mcg min Procainamide 2 g in 500 ml 4 g in 500 ml 0 g in 0 ml mg min Propofol mg kg min 200 mg in 20 ml 600 mg in 60 ml 0 mg in 0 ml mg kg min Propofol mcg kg min 200 mg in 20 ml 600 mg in 60 ml 0 mg in 0 ml mcg kg min Propranolol 15 mg in 500 ml 15 mg in 250 ml 0 mg in 0 ml mg hr Ranitidine 50 mg in 100...

Page 56: ...00 units in 500 ml 300 000 units in 200 ml 0 units in 0 ml units kg hr Vasopressin 150 units in 250 ml 200 units in 250 ml 0 units in 0 ml milli U min Vercuronium 5 mg in 5 ml 10 mg in 5 ml 0 mg in 0 ml mcg kg min Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 57: ...evF Name Concentration Dosage Units Caution Federal Law restricts this device to B Braun Medical Inc Carrollton TX USA 75006 sale by or on the order of a physician Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 58: ...heck Battery Level 5 1 5 2 Clear Total Infused 4 1 4 2 D Daisy Chaining 2 4 Data Keys 2 2 2 4 Dose Mode 2 1 4 1 6 1 Body Weight 6 2 Calculated Parameter 6 2 Change Medication 6 1 Change SETUP Parameters 6 3 Clear Medication 6 1 Concentration Screen 6 2 Data Screen 6 2 Drug Menus 6 1 Enter Data 6 3 Epidural 6 2 Exit Dose Mode 6 4 Titrate Dose 6 4 Titrate Rate 6 4 E Enter Key 2 2 2 4 Epidural Admini...

Page 59: ...ckout Switch 2 3 Piggyback Callback 3 3 5 1 5 5 Piggyback Fluids 3 2 Piggyback Indicator Key 2 2 Power Key 2 2 2 4 Profile Mode 5 1 7 1 Change Profile Parameters 7 1 Data Entry 7 1 Exit PROFILE Mode 7 3 Immediate Ramp Down 7 2 Restart PROFILE 7 2 Review Change PROFILE Data 7 2 Set Profile Parameters 7 1 Program Mode 5 1 7 4 Data Entry 7 4 Exit PROGRAM Mode 7 6 Restart PROGRAM 7 6 Review Change PRO...

Page 60: ...ule Next Run 5 1 5 9 Select MAX Pressure 5 1 5 10 Set Time Volume 5 1 5 11 Site Trend 5 1 5 12 Standard Mode 2 1 For Primary Fluids 3 2 For Piggyback Fluids 3 2 Sympathetic flow 3 3 T Titration 3 3 6 4 Artisan Technology Group Quality Instrumentation Guaranteed 888 88 SOURCE www artisantg com ...

Page 61: ...l service in house repair center WE BUY USED EQUIPMENT Sell your excess underutilized and idle used equipment We also offer credit for buy backs and trade ins www artisantg com WeBuyEquipment REMOTE INSPECTION Remotely inspect equipment before purchasing with our interactive website at www instraview com LOOKING FOR MORE INFORMATION Visit us on the web at www artisantg com for more information on ...

Reviews: