Horizon Nxt® with MedScan
Modular Infusion System 5 - 11 6/00
950520 Rev F
See “Occlusion Limit” to change the value.
Set PROFILE Parameters
Set PROFILE Parameters is described in Chapter 7 (Special Modes).
Set Time and Volume
Use this feature when an infusion rate is defined by time and
volume. The pump calculates the delivery rate.
1) Press
H
OLD
key
2) Press
M
ENU
key
3) Press
C
URSOR
keys to select “Set Time and Volume”
4) Press
E
NTER
key
1) Press
C
URSOR
keys to select “hr”
or “min”
2) Press
D
ATA
keys to enter the
correct hours and minutes
3) Press
C
URSOR
keys to select “ml”
4) Press
D
ATA
keys to enter the
volume
5) Press
C
URSOR
keys to select
“VALIDATE”
6) Press
E
NTER
key.
7) Press the
R
UN
key to start the infusion.
Set Time and Volume
Rate = 125.00 ml/hr
Time = 8 hr 00 min
Volume = 1000 ml
VALIDATE