Horizon Nxt® with MedScan
Modular Infusion System 5 - 12 6/00
950520 Rev F
Site Trend
Site Trend graphically displays the resistance to downstream fluid
flow for three intervals of time MAX, MED and MIN.
1) Press
M
ENU
key
2) Press the
C
URSOR
key to select
Alternate Menu
3) Press
E
NTER
key
4) Press
C
URSOR
keys to select Site
Trend
5) Press
E
NTER
key
Note:
Data is shown from right to left. The word "Now" reinforces
the most current data is represented on the right side of the
graph. The arrow’s position, whether pointing upward or
downward, indicates either an increase or a decrease in the
downstream fluid resistance.
To Change Graphical Resolution:
1) Press up and down
C
URSOR
key to change from MAX, MED or
MIN resolution
Note:
MAX is the default setting. MIN displays more data points
compressed along the horizontal axis for a representation of a long-
term trend. MAX displays less data points in the same area for a
short-term trend.
Resistance Factor (RF):
Note:
This feature needs to be activated by the Biomedical
Technician.
The Resistance Factor is a numeric value displayed and updated
during an infusion. "RF =" appears on either the Hold or Run Screen
when activated. The numerical value of RF appears after the words
"Site Trend" when the graph is displayed. Increasing RF values may
alert the clinician to a possible occlusion state due to kinks, clogged
filters, viscous fluids or infiltrations.
To Exit:
1) Press right
C
URSOR
key to select Exit
2) Press
E
NTER
key to exit
Exit from Menu
MAX Site Trend Exit
Now