Horizon Nxt® with MedScan
Modular Infusion System 5 - 9 6/00
950520 Rev F
Schedule Next Run
This feature lets the user delay an infusion from Standard, DOSE,
PROFILE or PROGRAM Mode for up to 24 hours.
Note:
The rate and volume information must be entered on either
the Primary or Piggyback channel and a set must be in the
pump before entering this feature.
1) Press
M
ENU
key
2) Press
C
URSOR
keys to select Schedule Next Run
3) Press
E
NTER
key
To Change the Time:
1) Press
C
URSOR
key to select
“Time is”, if necessary
2) Press
D
ATA
keys to set the
correct time (in military time)
To Change the Delay:
Note:
The time is displayed in military time. Use 00:00 to represent
midnight and 12:00 to represent noon.
Either:
Press
C
URSOR
keys to select “hr” or “min”
Press
D
ATA
keys to set the hours and minutes
or
Press
C
URSOR
keys to select “Start at”
Press
D
ATA
keys to change the time
1) Press
C
URSOR
keys to select “Validate”
2) Press
E
NTER
key
The information is displayed on the Information Screen, “dLY” is
shown, and the yellow Hold LED will blink. Press the
H
OLD
key to
cancel Schedule Next Run.
Note
: To prevent tampering of the Schedule, press the Panel Lock
Out Switch on the rear of the pump.
Schedule Next RUN
Time is 15:15p
Delay 2 hr 00 min
Start at 17:15p
VALIDATE 01 H 00 M DELAY