Horizon Nxt®
Modular Infusion System 7 - 4 6/00
950520 Rev F
PROGRAM Mode
This mode lets the user program up to nine different rates and volumes. The
pump will deliver the fluid sequentially. Fluid delivery can be customized
for each patient.
If the pump was turned off while in the PROGRAM Mode, this screen will
appear:
1) Press
C
URSOR
keys to select the
appropriate message
2) Press
E
NTER
key
The pump will return to Program Hold state.
1) Press
M
ENU
key
2) Press
C
URSOR
keys to select "PROGRAM Mode"
3) Press
E
NTER
key
Data Entry:
1) Press
C
URSOR
keys to select Rate or
Volume
2) Press
D
ATA
keys to change the values
3) Press
C
URSOR
keys to select the next
period
4) Repeat until all information is entered
(up to 9 periods)
Note:
VALIDATE cannot be selected when 0.0 data is present in the
middle of a Program, for example:
NOT ALLOWED
PROGRAM Operation
was Interrupted
Resume current PROGRAM
Restart PROGRAM
Exit PROGRAM Mode
Pd. Rate Volume
1 0.0 0.0 ZERO
2 0.0 0.0
3 0.0 0.0
VALIDATE TOTAL = 0.0 ml
Pd. Rate Volume
2 0.0 0.0
3 30.0 150.0 ZERO
VALIDATE TOTAL = 0.0 ml
1 50.0 35.0