Horizon Nxt® with MedScan
Modular Infusion System 5 - 6 6/00
950520 Rev F
1) Press right
C
URSOR
keys to select
OK
2) Press
E
NTER
key
To Enter Data:
1) Press
C
URSOR
keys to select "
q(x)
hrs and min"
2) Press the
D
ATA
keys to set the
time interval between infusion.
(For example:
q
6 hrs 00 min)
Note:
The time interval can be 1 to 24 hours. There must be at least
15 minutes between each run.
1) Press down
C
URSOR
key to select "# of Doses"
2) Press
D
ATA
keys to set the total number of doses to be infused.
Note:
The total number of doses may not exceed 40.
1) Press down
C
URSOR
key to select "Dose Rate"
2) Press the
D
ATA
keys to set the infusion rate in ml/hr.
1) Press
C
URSOR
keys to select "Dose Vol"
2) Press the
D
ATA
keys to set the volume of each dose (for example:
4 doses, each 50 ml)
Note:
The Dose Volume of each dose has to be greater than 1 ml
but not more than 2000 ml.
3) Press the
E
NTER
key
To change the time:
1) Press up
C
URSOR
key to select
"Present Time is"
2) Press
D
ATA
keys to set the correct
time
Note:
The time is displayed in military time. Use 00:00 to represent
This feature allows
for periods of no
infusion. Ensure IV
patency prior to start
Cancel OK
Dose q(x) 01 hr 00 min
# Doses = 4
Dose Rate = 200.0 ml/hr
Dose Vol = 100 ml
Press ENTER when done
Present Time is
09:37a
Start dose at
10:35a
VALIDATE
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com