121
(2)
The input voltage is displayed on the screen. Input this voltage into the recorder.
(3)
Adjust the zero point
。
Exampl
e:
To adjust the
±
2V range
•
Input 0V by the DC voltage current generator.
(4) Wait about 5 seconds after inputting the zero voltage, and then press [Go] button.
(5) Adjust the span.
Example: Adjustment of the ±2 V range
Input a voltage of +2 V with the DC voltage current generator.
(6) Wait about 5 seconds after inputting the span voltage, and then press [Go] button.
(7) After the adjustment of the span, the screen will return to the calibration screen for all
ranges.
(8) Repeat steps (1) to (6) if adjustment is needed for other ranges.