64
CONDITIONING FLUID NOT DETECTED
When the conditioning cycle has been initiated, the analyzer checks for proper
aspiration of the conditioning solution using the sample sensor. If the solution is
not detected in the proper time, this error message will be displayed.
Possible causes and remedies:
•
Ensure that a sufficient volume of conditioning solution is presented for
aspiration and no air bubbles are present as solution is aspirated.
•
Check to ensure that the solution is being properly aspirated
through the sampling mechanism. The sample path must be free of
leaks or clots.
•
Verify that sample sensor is correctly plugged in and perform the
SAMPLE SENSOR TEST
to verify its proper operation.
•
Check pump tubing and replace as necessary to ensure correct
pump rate.
INTERFACE ERROR
This message is displayed only during the
INTERFACE TEST
. Pins 2 and 3
must be shorted together to perform the
INTERFACE TEST
. Make
sure
that the pins are not shorted to chassis ground.
Possible cause and remedy:
•
Replace SBC Card.
PAPER JAM OR PRINTER DEFECT
This message is displayed when the printer head is jammed and tries to print. It
is a temporary message that is displayed for 2 seconds and then sample results
are displayed. To clear paper jam remove printer from the analyzer and
remove jammed paper. Replace printer and retry. The printer can be removed
by grasping the rear edge of the printer assembly and pulling it to the front. The
printer can be removed with the Analyzer power on. To free up jammed paper,
turn the small gears on the left side of the printer module.