46
DEC CED
Provides one additional digit of resolution for each
measured parameter for blood and serum samples.
QC and Standard samples are always displayed with
high resolution. Urine samples are always displayed in
low resolution.
ECO OCE
Allows the operator to run the instrument in the
economy mode. The instrument will enter the
STANDBY MODE
. This occurs at the time of the
next 2-point calibration when no samples have been
analyzed since the last 2-point calibration.
FIF
IFI
Allows printing of electrode voltages during printout
of measurements and calibrations.
(not available for 9140)
LEM MEL
This service code will print the last 20 error messages
(9180/9181 only)
with the date and time of occurrence.
MGL LGM
This service code is available only in the model 9140/
9180/9181 analyzer and enables Ca
++
values to be
displayed in mg/dL. The normal mode displays values in
mmol/L. QC and Standard results are always displayed
in mmol/L.
NOB BON
This service code disables the beep.
(9180 /9181only)
QCC CCQ
When this code is enabled, QC values will be reported in
direct ISE measurement instead of flame-equivalent
values. In this mode, correlation factors are NOT
applied to the measured values.
SFC
This code allows the input of a percentage for the ISE
(9180/9181 only)
SnapPak
TM
counter.
SLC
CLS
When this code is set, the current setting of the leasing
(9180/9181 only)
counter, which cannot be reset by the operator, is
printed on the calibration report.
SSC CSS
This code toggles needle height to accommodate smaller
(9181 only)
sample cups. In the default setting, the analyzer works
with 2.0 mL cups. To select 0.25 mL sample cups,
SSC
has to be activated. To return to the default
setting,
CSS
has to be activated.