NO SAMPLE
This message is displayed when any of the following conditions occur:
1.
During sampling, either a bubble in the sample is detected or thesample size
is smaller than the minimum sample required.
2.
The sample door is opened and no sample is fed into the analyzer.
3.
Sample door is not closed within 20 seconds after sample is detected.
Possible causes and remedies:
•
Check whether the sample sensor is plugged in and perform
TEST SAMPLE SENSOR
to verify that it operates correctly.
•
Check sample draw and look for clots or leaks in the sampling system.
CLEANING FLUID NOT DETECTED
When the cleaning cycle has been initiated, the analyzer checks for proper
aspiration of the cleaning solution using the sample sensor. If solution is not
detected in the proper time, above error message will be displayed.
Possible causes and remedies:
•
Ensure that a sufficient volume of cleaning solution is presented for
aspiration and no air bubbles are present as solution is aspirated.
•
Check to ensure that the solution is being aspirated through the sampling
mechanism correctly. The sample path must be free of leaks or clots.
•
Verify that sample sensor is properly plugged in and perform the
SAMPLE SENSOR TEST
to verify the correct operation of the
sample sensor.
•
Check pump tubing and replace as necessary to ensure a correct pump
rate.