Avidity Science, LLC
Pico Water Purification System Operation Manual
Maintenance
27
18.
Check for leaks from the UV lamp connection at the reservoir.
19.
Take one of these actions.
20.
Drain the reservoir until the water level is below the UV lamp.
21.
Disconnect the power supply.
22.
Remove the right front panel.
23.
Disconnect the UV lamp from its power connector.
24.
While holding the socket on the outside of the tank, insert your hand into the reservoir and loosen
and remove the nut that secures the UV lamp.
25.
Again, while holding the lamp socket in place, secure the nut to the socket from inside the reser-
voir.
26.
27.
Secure the cap to the top of the reservoir.
28.
Secure the right cover with the four hex head screws and replace the left cover.
29.
Secure the left cover into position.
If a leak is...
Then...
Not present,
Present,
go to step 20.
Summary of Contents for Pico PICO10T2
Page 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Page 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Page 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Page 38: ......
Page 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...