Avidity Science, LLC
Pico Water Purification System Operation Manual
Maintenance
21
16.
When Change Complete appears, touch
17.
Replace the left cover.
Replace Pre-Treatment / RO Module
The Pre-Treatment and RO module are combined in one canister. Once the Pre-Treatment/RO Module
has been replaced it will have to be rinsed.
IMPORTANT:
If a DI cartridge is required for Type 2 water production, the bypass tube will need to be con-
nected between the green RO water outlet tube and the white reservoir inlet tube.
Follow this procedure.
1.
Remove the left cover from the Pico unit.
1.
On the control display, press
.
2.
Press .
3.
Select
TC021/TC022
. Pressure is relieved from the module.
4.
Disconnect the Blue inlet tubing from the top of the module
5.
Disconnect the Red permeate and Black concentrate tubing from the module.
6.
Lift the module from the mounting slots on the back wall of the unit.
7.
Remove the new Pre-Treatment and RO module from its packaging.
8.
Insert the hooks of the canister into the slots located on the right side.
CAUTION!
De-pressurize the system before disconnecting the Pre-Treatment
and RO module. Failure to do so can result in personal injury and equipment
damage.
Summary of Contents for Pico PICO10T2
Page 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Page 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Page 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Page 38: ......
Page 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...