Pico Water Purification System Operation Manual
Avidity Science, LLC
30
Recording Monthly Operating Data
Percent Rejection
A percent rejection setting of 95 percent
is used to rate the performance of a reverse osmosis water
machine. A noticeable decrease in the percent of rejection is an indication that a membrane should be
replaced. Membrane fouling and/or scaling, hydraulic plugging, membrane hydrolysis (degradation), and
mechanical failures can plug up the membrane and the percent of rejection can increase.
Membrane fouling and/or scaling
is the presence of organic or biological foulants and/or mineral scale buildup
directly on the membrane surface.
Hydraulic plugging
is the presence of material (foulant, scale or particles) that obstructs the flow-channel
spacing between the membrane leaves of a spiral wound membrane.
Table 5 lists some of the issues that can be identified by a change in the percent rejection rate. In addition
to these issues, a drop in the percent of rejection may be a sign of an improper pH level, a sign the percent
recovery rate is set too high, or a sign the composition of the feed water has changed.
Below are the formulas required to calculate the Percent Rejection.
Table 5. The Effect of Various Problems on Percent Rejection.
Problem
Percent of Rejection
Rate of Change
Biofouling
initial increase, decrease after cleaning
gradual
Chemical scale
decrease, especially in last membrane
rapid
Colloidal fouling
initial slight decrease, much greater later
gradual
Aluminum sulfate scale
little to no change
gradual
Leaking o-rings
decrease
immediate
RO unit pump pressure too low
decrease
immediate
AverageConductivity
FeedWaterConductivity ConcentrateConductivity
+
2
-----------------------------------------------------------------------------------------------------------------------------------------------------
=
PercentRejection
AverageConductivity PermeateConductivity
–
AverageConductivity
------------------------------------------------------------------------------------------------------------------------------------
100
=
Summary of Contents for Pico PICO10T2
Page 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Page 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Page 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Page 38: ......
Page 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...