Pico Water Purification System Operation Manual
Avidity Science, LLC
10
Installation and Setup
2.
Connect one end of the bypass tubing (
A
) to the open end of the
green
RO water outlet tube (
B
)
and the open end of the
white
reservoir inlet tube (
C
) as shown in Figure 6. Use the provided
elbow fittings.
Figure 6. Bypass tubing connected.
3.
On the control display, press
.
4.
Press .
5.
Press
. The inlet valve will open and begin flushing the unit.
6.
When asked if RO module is installed, touch
7.
Allow the pressure to adjust and stabilize for approximately 3 minutes and then take one of these
actions.
8.
Turn off the water to the Pico unit.
If the pressure is
Then,...
Less than 0.5 bar
Greater than 0.6 bar
Press
OK
and got to step 13
A
B
C
Summary of Contents for Pico PICO10T2
Page 1: ...Pico Water Purification System Operation Manual OP 000100 Revision D August 2018...
Page 22: ...Pico Water Purification System Operation Manual Avidity Science LLC 16 Installation and Setup...
Page 34: ...Pico Water Purification System Operation Manual Avidity Science LLC 28 Maintenance...
Page 38: ......
Page 40: ...Pico Water Purification System Operation Manual Avidity Science LLC 34 Replacement Parts...