Pico Water Purification System Operation Manual
Avidity Science, LLC
24
Maintenance
20.
When operation is complete, press firmly at the center of the timer label sticker until a thin red line
appears. See Figure 15.
Figure 15. Red line shows RO module timer is activated.
21.
If a DI cartridge is not used, replace and secure the left front cover of the Pico unit.
IMPORTANT:
If applicable, replace the DI cartridge for Type 2 water production.
Replace DI Cartridge (for Type 2 water quality only)
Follow this procedure.
1.
Remove the left cover from the Pico unit.
1.
On the control display, press
2.
Press
3.
Select
TC023
for the DI cartridge. The pressure is relieved from the cartridge.
4.
Disconnect the green tubing from the bottom of the cartridge.
5.
Disconnect the white tubing from the top of the cartridge.
6.
Lift the cartridge from the mounting slots on the back wall of the unit.
CAUTION!
De-pressurize the system before disconnecting the DI cartridge.
Failure to do so can result in personal injury and equipment damage.