QSG_F670_363.00_EN_01_2021-02-09
24/42
Direct mode
– High range + External Dump option
This configuration is used for direct mode at maximum pressure 2 MPa (290 PSI).
Connection
Option / description
Air supply to
1
Connection of the air supply to the filter input (0.6 MPa (87 PSI))
Regulator air supply
to
2
Connection of the regulator air supply (2 MPa (290 PSI))
3
to ATM
Output to be left to the atmosphere (exhaust of electronic regulator)
7
to
4
Connection of the Test output to the Y valve input (External Dump option)
5
to
6
Connection of the reference output to the reference part
8
to ATM
Output Y2 to be left to the atmosphere (exhaust of the part under test and the
reference part)
9
to
10
Connection of the Y1 output to the part 1 under test (External Dump option)