QSG_F670_363.00_EN_01_2021-02-09
36/42
User adjustments
OPTIONS OF THE MENUS
Different menus are accessible on the
MAIN MENU
screen.
For more information, refer to the Reference Manual.
SPECIAL CYCLE MENU menu
Use this menu to carry out specific procedures necessary to ensure the proper operation of
measurement cycles (for example, adjustment of pressure regulator).
Some parameters are displayed when specific functions are activated.
TO START SPECIAL CYCLES...
1. On the
SPECIAL CYCLE MENU
screen, select a cycle, and press
to validate.
2. Press
Start
to execute the cycle.
3. To stop the current cycle press
Reset
.
Label
Special cycle
Description of the cycle
none
None
No special cycle selected
Regulator adjust.
Regulator adjustment
Adjustment of regulator in front panel
Inf Fill
Infinite fill
Pressurise the part with an infinite fill time
Piezo auto zero
Piezo auto zero
Auto zero cycle on the piezo sensor
AUTO VOL
Automation volume
Volume calculation for automatic program
selection
Label
Special cycle
Description of the cycle
Part. Regulator Adj.
Part regulator
adjustment
Regulator adjustment for indirect mode
Custom Unit Learn
Custom unit learn
Define unit on a master leak
Custom Unit Check
Custom unit check
Check correct unit learning on a master leak
Check+Lrn Cust. Unit
Custom unit check and
learn
Check correct unit learning and relearn if
necessary
Sd Prt PASS Learn
Sealed part pass learn
Mandatory cycle for sealed part test mode
Sd Prt FAIL Learn
Sealed part fail learn
Reject level adjustment for sealed part test
mode
LEAK OFFSET LEARN
Leak offset learn
Special cycles to determine offset parameters
VOL. LEARN
Offset and volume learn
Special cycles to determine offset and volume
parameters (Flow unit)
N START
Number start cycle
Launch a determined quantity of cycles