CALIBRATION
93
9. DC Hipot, 3500
m
A current range
Press the
Ú
key, the program will advance to the DC 3500
m
A range calibration of
hipot test. The display will show:
DC 3500
m
A, 100K
W
<
T E S T
>
t o C a l i b r a t e
Connect a resistor about 100K
W
/10W in series with a DC standard Ammeter which
can measure up to 3000
m
A to the output leads. The Ammeter should be connected
to the return lead, then press the TEST button, the program will automatically
generate an output of about 300V/3000
m
A and the display will show:
Current =
m
A
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit
m
A, and then press
the ENTER key to store the standard value of DC 3500
m
A range of hipot test and
the display will show:
DC 3500
m
A, 100K
W
OK
<
T E S T
>
t o C a l i b r a t e
10. DC Hipot, 350
m
A current range
Press the
Ú
key, the program will advance to the setting and reading of DC 320
m
A
range calibration of the hipot test. The display will show:
DC 350
m
A, 1M
W
<
T E S T
>
t o C a l i b r a t e
Connect a resistor about 1M
W
/0.25W in series with a DC standard Ammeter which
can measure up to 300
m
A to the output leads. Connect the Ammeter to the return
leads, then press the TEST button, the program will automatically generate an
output of about 300V/300
m
A and the display will show:
Current =
m
A
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit
m
A, and then press
the ENTER key to store the standard value of DC 350
m
A range of the hipot test and
the display will show:
DC 350
m
A, 1M
W
OK
<
T E S T
>
t o C a l i b r a t e