SPECIFICATIONS
13
Model 7504SA, and 7500DT
Functional Specifications
INPUT
Voltage
100/115/200/230 VAC
±
10
%
, Single Phase, User selection
Frequency
47 - 63 Hz
Fuse
15 Amp 250V fast blow internal
DIELECTRIC WITHSTAND TEST MODE
Output
Rating
5 KV @ 100 mA AC,
6 KV @ 10 mA DC
Output Adjustment
Range:
0
- 5
KV
AC, 0
- 6
KV
DC
Resolution: 1 volt/step
Accuracy:
±
(2
%
of setting
+
5 volts)
Ramp-HI
12mA peak maximum, ON/OFF selectable
Charge-LO
Range:
0.0 - 350.0
m
A DC or Auto set
HI-Limit
AC Range:
0.00
-
99.00
mA
Resolution: 0.01 mA/step
DC Range:
0 - 9999
m
A
Resolution: 1
m
A/step
Accuracy: AC or DC
±
(2
%
of setting
+
2 counts)
LO-Limit
AC Range:
0.000
-
9.999
mA
Resolution: 0.001 mA/step
DC Range:
0.0 - 999.9
m
A
Resolution: 0.1
m
A/step
Accuracy: AC or DC
±
(2
%
of setting
+
2 counts)
Arc
Detection
Range:
1
-
9
Failure
Detector
Audible
and
Visual
Voltage
Display
Range:
0.00 - 6.00 KV Full Scale
Resolution: 10 volt/step
Accuracy:
±
(2
%
of reading
+
2 counts)