OPERATION
58
At the beginning of the Ground Bond test, before the meter readings are taken, the
display will show:
GND Dwell XXX.X s
MXX-X - - . - - A - - - m
W
During the Ground Bond test, when the values are being updated in real time, the
display will show:
GND Dwell XXX.X s
MXX-X XX.XX A XXX m
W
If the DUT failed the HI-Limit of the Ground Bond test and the ground continuity
resistance is within the metering range, the display will show:
GND HI-Limit XXX.X s
MXX-X XX.XX A XXX m
W
If the DUT failed the HI-Limit of the Ground Bond test and the ground continuity
resistance is beyond the metering range, the display will show:
GND HI-Limit XXX.X s
MXX-X 0.00 A
>
600 m
W
If the DUT resistance falls below the LO-Limit of the Ground Bond test, the display
will show:
GND LO-Limit XXX.X s
MXX-X XX.XX A XXX m
W
When the DUT passed the Ground Bond test when the test process is complete, the
display will show:
GND Pass XXX.X s
MXX-X XX.XX A XXX m
W
5.2 The followings are the displayed messages for the AC Withstand Voltage test:
If the test in process is Aborted with the RESET button or remote control, the
display will show:
ACW Abort XXX.X s
MXX-X X.XX KV XX.XX mA
If the test in process is Aborted with the RESET button or remote control before the