SPECIFICATIONS
15
INSULATION RESISTANCE TEST MODE
Output
Voltage
Range:
100 - 1000 Volts DC
Resolution: 1 volt/step
Accuracy:
±
(2
%
of reading
+
2 volts)
Short
Circuit
Current Maximum: 12mA
peak
Voltage
Display
Range:
0
-
1000
V
Resolution: 1 volt/step
Accuracy:
±
(2
%
of reading
+
2 counts)
Resistance
Display
Range:
1 - 9999 M
W
(4 Digit, Auto Ranging)
Resolution:
500VDC
1000VDC
M
W
M
W
M
W
0.001 1.000
-
5.388
1.000
-
9.999
0.01
1.40
-
53.88
2.80
-
99.99
0.1
14.0
-
538.8
28.0
-
999.9
1
104
-
9999
280
-
9999
Accuracy:
±
(2
%
of reading
+
2 counts) at test voltage
500 - 1000V and 1 - 1000 M
W
±
(8
%
of reading
+
2 counts) at test voltage
500 - 1000V and 1000 - 9999 M
W
±
(8
%
of reading
+
2 counts) at test voltage
100 - 500V and 0 - 1000 M
W
Charge-LO
Range:
0.000 - 3.500
m
A or Auto Set
HI-Limit
Range:
0 - 9999 M
W
(0 = Off)
LO-Limit
Range:
1 - 9999 M
W
Delay
Timer
Range:
0, 0.5 - 999.9 sec
(0 = Constant)
Resolution: 0.1 sec/step
Accuracy:
±
(0.1% + 0.05 sec)
Ground
Fault
Interrupt
GFI Trip Current: 450
m
A max (AC or DC)
HV Shut Down Speed: < 1 ms