OPERATION
45
2.3.2 Charge-LO
Advance the menu to the Charge-LO parameter. The display will show:
Charge-LO = X.XXX
m
A
<
TEST
>
to Auto Set
The Charge-LO function is used to check if the test cables are connected properly at
the beginning of a test. A capacitive DUT will draw charging current on the
Insulation Resistance test when the Output is activated. If the charging current was
lower then the setting, the test cables may not be connected properly.
This instrument can set the Charge-LO value manually or automatically. To set the
Charge-LO value manually, use the numeric keys to enter the Charge-LO current
setting and then press the ENTER key . The program will store the Charge-LO
setting and advance to the HI-Limit parameter. The setting range of Charge-LO is
from 0.000 to 3.500
m
A. The unit is
m
A and 0.001
m
A per step.
To use Auto Set, be sure that the test parameter of Output Voltage has been set to
the values that will be used for the final test and connect the test cables and/or test
fixture between the instrument and DUT. If the scanner is to be used then the
scanner channel must also be set. Then press the TEST button. The instrument will
apply the voltage that has been entered for this memory-step selection.
Please be aware that the program will activate high voltage on
the output connector while the Test button is pressed.
The program will read the charging current of DUT and set the Charge-LO current
at approximately one half (1/2) of the reading. The display will show:
Charge-LO = X.XXX
m
A
<
TEST
>
to Auto Set
The value showing on the display is the Charge-LO setting and is not the reading of
the charging current of DUT.
Then press the
Ù
or
Ú
key to advance the program to another test parameter or press
the EXIT key to exit from the setting mode to the operation mode.
2.3.3 HI-Limit Current setting
Advance the menu to the HI-Limit parameter. The display will show:
HI-Limit = XXXX M
W
Range : 1 - 9999
WARNING