OPERATION
37
2.1.3 LO-Limit Current setting
Advance the menu to the LO-Limit parameter. The display will show:
LO-Limit = X.XXX mA
Range : 0.000 - 9.999
Use the Numeric keys to enter the LO-Limit setting, then press the ENTER key.
The program will store the LO-Limit setting and advance to the Ramp Time
parameter automatically. The unit is mA and 0.001 mA per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
If the LO-Limit is set to
, the LO-Limit is disabled.
2.1.4 Ramp Time setting
Advance the menu to the Ramp Time parameter. The display will show:
Ramp Time = XXX.X s
Range : 0.1 - 999.9
Use the Numeric keys to enter the Ramp Time setting, then press the ENTER key.
The program will store the Ramp Time setting and advance to the Dwell Time
setting automatically. The unit is second and 0.1 second per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
2.1.5 Dwell Time setting
Advance the menu to the Dwell Time parameter. The display will show:
Dwell Time = XXX.X s
Range : 0.3 - 999.9 0 = Constant
Use the Numeric keys to enter the Dwell Time setting, then press the ENTER key.
The program will store the Dwell Time setting and advance to the Frequency
selection parameter automatically. The unit is second and 0.1 second per step. The
display will show the elapsed time during the testing.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
If the Dwell Time is set to
, the timer will continue to count to the maximum test
time then reset to 0 and start over automatically. The test will continue until a reset
is executed or a failure occurs.