OPERATION
47
2.3.6 Scanner Channel setting
Advance the menu to the Scanner Set parameter. The display will show:
Scanner Set CH = 1 - 16
XXXXXXXXXXXXXXXX
Note: X = H, O or L.
Use the Numeric keys to enter the status of High, Open or Low for each scanner
channel, then press the ENTER key. The program will store the Scanner Channel
status and advance to the Connection setting automatically. The following table is
the cross reference of the numeric keys, which are used for the Scanner channel
setting.
Channel
Status
Numeric
Key
Description
H (High)
1
Connected to High Voltage Output
L (Low)
0
Connected to Return
O (Open)
·
Not connected to High Voltage or
Return
The Channel setting will start from Channel 1 on the left end and stop at Channel 16
on the right end. The program will set any unassigned channel(s) to OPEN
automatically after pressing the ENTER key. Therefore it is necessary to set only
the channels that are required for the test.
If the external Scanner is not connected, this setting is not necessary.
2.3.7 Step Connection setting
Advance the menu to the Connection parameter. The display will show:
Connect = ON
<
ENTER
>
to Select
or
Connect =
OFF
<
ENTER
>
to Select
Use the ENTER key to select the Step Connection mode, then press the
Ù
or
Ú
key
to advance the program to another test parameter or press the EXIT key to exit
from the setting mode to the operation mode.
If Connect is set to ON, the next step in the sequence will be executed. If Connect is
set to OFF, the test sequence will stop at this step.
If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
2.4 Ground Bond test setting model 7504SA only
Press the GND key and the display will show:
GND Set XXX.X s
MXX-X XX.XX A XXX m
W
Note: X = the numeric