![Associated Research HypotULTRA II 7500DT Operation And Service Manual Download Page 20](http://html1.mh-extra.com/html/associated-research/hypotultra-ii-7500dt/hypotultra-ii-7500dt_operation-and-service-manual_2989479020.webp)
SPECIFICATIONS
16
GROUND BOND TEST MODE Model 7504SA only
Output
Voltage
Range:
3.00
-
8.00
Volts
AC
(Open
Circuit
Limit)
Resolution: 0.01
volt/step
Accuracy:
±
(2 % of S 0.03V ) O.C. Condition
Output
Frequency
Range:
60
or
50
Hz,
User
Selection
Accuracy:
±
1%
Output
Current
Range:
3.00
-
30.00
Amps
AC
Resolution: 0.01
Amp/step
Accuracy :
±
(2 % of S 0.02 A)
Current
Display
Range:
0.00
-
30.00
Amps
Resolution: 0.01
Amp/step
Accuracy:
±
(3 % of R 0.03 A)
Resistance
Display
Range:
0 - 600 m
W
Resolution:
1 m
W
/step
Accuracy:
±
(2 % of R 2 m
W
)
HI-Limit
Range:
0 - 600 m
W
for 3 - 10 A
0 - 150 m
W
for 3 - 30 A
Resolution:
1 m
W
/step
Accuracy:
±
(2 % of S 2 m
W
)
LO-Limit
Range:
0 - 600 m
W
for 3 - 10 A
0 - 150 m
W
for 3 - 30 A
Resolution:
1 m
W
/step
Accuracy:
±
(2 % of S 2 m
W
)
Dwell
Timer
Range:
0, 0.5 - 999.9 sec (0 = Constant)
Resolution: 0.1
sec/step
Accuracy:
±
(0.1% + 0.05 sec)
Milliohm
Offset
Max. Offset
Capability:
200 m
W
Resolution:
1 m
W
/ step
Accuracy:
±
(2 % of S 2 m
W
)