OPERATION
50
Use the ENTER key to select the Output Frequency, then press the
Ù
or
Ú
key to
advance the program to another test parameter setting or press the EXIT key to exit
from the setting mode to the operation mode.
2.4.7 Scanner Channel setting
Advance the menu to the Scanner Set parameter. The display will show:
Scanner CH = X
Range: 1-16 0 = OFF
Use the Numeric keys to enter the scanner channel number, then press the ENTER
key. The program will store the scanner channel number and advance to the Offset
parameter automatically.
If the external Scanner is not connected, it is not necessary to enter this parameter.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
To operate from the front panel output connectors with the Scanner option installed,
the Scanner Channel must be set to
to receive accurate readings.
2.4.8 Offset setting
Advance the menu to the Offset parameter. The display will show:
Offset = XXX m
W
<
TEST
>
to Auto Set
This instrument can set the m
W
offset value by two different methods. One is the
Manual Offset and the other is Auto Offset.
For Manual Offset, use the Numeric keys to enter the m
W
Offset value and then
press the ENTER key. The program will store the m
W
Offset setting and advance to
the Connect setting. The range of m
W
Offset is from 0 to 200 m
W
. The unit is m
W
and 1m
W
per step.
For Auto Offset, setup the test parameter of Output Voltage, Current and Frequency
and connect the test cables and/or test fixture to the instrument first. Then short
circuit the ends of the test cables. If the scanner is being used, the scanner channel
has to be set. Then press the TEST button. The program will activate the test
current and frequency that has been set for this memory-step. The display will show:
Offset = XXX m
W
<
TEST
>
to Auto Set