expansion chamber at the top of the thermometer. Note that over filling the
expansion chamber will break the thermometer. Tap the thermometer gently on
the surface of a large rubber stopper in an upright position allowing the gas
separating the column to rise above the column. Allow the thermometer to cool
slowly in an upright position.
Cooling Method:
Prepare a solution of shaved ice and salt or CO
2
(Dry-Ice)
and alcohol. Place the thermometer bulb only in the solution. Keep the
thermometer upright. Allow the liquid column to retreat into the bulb, swing the
thermometer (Bulb Down) in the arc releasing the entrapped gas permitting it to
escape above the column. Allow the thermometer to warm slowly in an upright
position.
LISTING:
CUL (100-120V~), EN61010-1 (230V~), UL (100-120V~)
SYMBOL DESCRIPTION:
~
AC Alternating Current
I
ON (Power Supply)
Caution
Hot Surface
Protective Earth
O
OFF (Power Supply)
Fuse
~ ~~
4
5. Disconnect power before cleaning or servicing. Servicing should be done by
a qualified technician only.
6. Risk of electric shock. Do not open.
7. Do not use in an explosive atmosphere.
8. Do not touch white block surface when incubator is turned on. Surface is
hot.
9. Close cover during incubation period.
10.Replace fuses with those of the same type and rating. (Applicable to Model
21006 only.) Fuse size: 5 mm diameter x 20 mm length; Fuse rating: 1 AMP,
250 Volt.
CLEANING:
Unplug STERRAD Incubator. Wipe down with a lightly dampened cloth of
sodium hypochlorite, ensuring that no liquid gets in the incubator. Dry with a
clean cloth. Contact ASP before using any cleaning solution not recommended
by the manufacturer.
PERFORMANCE CHARACTERISTICS AND SPECIFICATIONS:
SERVICE RATING
TUBE CAPACITY
HEATING TIME TO 58°C
Continuous
18
1 hour
MODEL
VOLTAGE
AMPERAGE
POWER
FREQUENCY
TEMP SETPOINT
21005
100-120 ±10% V~
0.4A
33-48W
50/60 Hz
58 ±2°C
21006
230 ±10% V~
0.2A
44W
50/60 Hz
58 ±2°C
ENVIRONMENTAL CONDITIONS:
Operating:
15° - 30°C; 10 – 85% relative humidity at 25°C, non-condensing
Installation Category II (overvoltage) in accordance with IEC 664
Pollution Degree 2 in accordance with IEC 664
Tested to an altitude of 2,000 meters
Storage:
–40°– 70°C; 10 – 85% relative humidity
THERMOMETER:
30° - 65°C, INDICATION ACCURACY ±0.5°C
Instructions For Reuniting Separated Fluid Column of Your EVER-SAFE™
Thermometer.
IMPORTANT: Store Thermometers in an upright position only. Handle
Instruments With Care. Wear Safety Glasses and Gloves Before Proceeding.
Heating Method:
Heat the thermometer bulb in an upright position away from
your face in warm liquid, air, or over a soft flame sufficient to allow the liquid
column to rise slowly until the separated portion of the column enters the
3
LC-21005-001B(2).qxd 9/7/04 9:47 PM Page 4
Summary of Contents for 21005
Page 21: ...40 39 LC 21005 001B 2 qxd 9 7 04 9 47 PM Page 40...
Page 22: ...42 41 J LC 21005 001B 2 qxd 9 7 04 9 47 PM Page 42...
Page 23: ...44 43 PL LC 21005 001B 2 qxd 9 7 04 9 47 PM Page 44...
Page 24: ...46 45 LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 46...
Page 25: ...48 47 H LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 48...
Page 26: ...50 49 SK LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 50...
Page 27: ...52 51 LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 52...
Page 28: ...54 53 CZ LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 54...
Page 29: ...55 LC 21005 001B 2 qxd 9 7 04 9 48 PM Page 56...