36
6
TROUBLE SHOOTING
Display
Possible lamp
(KL)
Possible cause
Solution
By
whom
E006
KL blasting gun
glows red
Short in blasting gun or
control cable.
Replace the blasting gun or
the control cable.
Operator
E007
KL com-
pressed air
glows red
Fault in the compressed
air regulator valve.
Call Customer Service.
Operator
E008
KL blasting gun
glows orange
The trigger of the blas-
ting gun was activated
during the switch-on
process or the releasing
of the emergency stop
key.
Release the trigger of the
blasting gun.
Operator
Display
Possible cause
Solution
By
whom
No display
in spite
of power
switch in
position 1
Main plug not connected
to the socket.
Insert the main plug into a socket.
Operator
No com-
pressed air
jet despite
the trig-
ger being
drawn
Fuse F1 blown.
Remove the side panel and unlock the fuse
F1 by pressing on it.
Operator
Compressed air supply
has too little pressure.
Check pressure level.
Operator
Jet pressure is set too low.
Set the jet pressure to a higher level.
Operator
Power supply has been
interrupted.
Check power supply. Indicator lamp „De-
vice on“ must glow green.
Operator
Emergency-stop button
has been pressed.
Release emergency-stop button by tur-
ning. Indicator lamp „Device on “must glow
green.
Operator
Control cable not connec-
ted properly.
Check connection between control cable
and the blasting gun and between the con-
trol cable and the device.
Operator
Control cable is defective.
Replace the cable
Operator