6 - 40
-Turn the lift ON by pushing on the red button and verify if the LCD
screen is working properly.
-Test the “up and down” alternate button.
Refer to section “INSTALLING THE TOP CABIN”, page 6-26
FROM SOFTWARE 2.2.X AND HIGHER, IF THE REPLACEMENT
BOARD ALREADY HAS DATA INSIDE ITS MEMORY, AT THE
STARTUP, THE SYSTEM WILL ASK YOU TO CONFIRM THE SE-
RIAL NUMBER OF THE LIFT.
NOTE
-The serial number of the lift is located on the product name plate on
the top cabin.
-Select the serial number of the lift and press the
button.
-Confirm your choice by selecting “yes” and pressing on the
button
.
Note that “back” can be used to return to the serial number
selection window.
001-15697 rev
. 5
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...