6 - 35
-Disconnect ECS connector (if applicable) (1)
-Disconnect the 4 pin connector (Motor encoder) (2)
-Disconnect the 12 pins connector (3)
-Disconnect green wire (ESD discharged) (4)
INSTALLING THE MAIN BOARD
-Slide the main board inside the plastic support.
-Fix the board in place with the two M5 screws.
DO NOT OVERTIGHTEN THE SCREWS TO AVOID
ANY DAMAGE TO THE MAIN BOARD.
MAKE SURE THE BOARD IS CORRECTLY INSTALLED ON ITS
SUPPORT.
ALWAYS ORDER MAIN BOARD FROM THE MANUFACTURER
TO AVOID LOSING INFORMATION (E.G. CALIBRATION DATA,
SERIAL NUMBER, HISTORICAL DATA, PREVENTIVE
MAINTENANCE SCHEDULE).
NOTE
NOTE
WARNING
1
2
4
3
FOR MS2 PLUS, ONLY REPLACE BOARDS WITH GENUINE
PARTS. NEVER SWAP BOARDS BETWEEN LIFTS.
WARNING
001-15697 rev
. 5
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...