CUSTOMIZING
-To access the Customizing section, press and hold the
button
for 3 seconds.
-Select “customizing” then press
to confirm.
-Press Up/Down button to scroll through the Customizing menu.
-Press
button to enter and accept value.
Parameters available in the customizing menu
Parameter
Description
Settings (default in bold character)
Motor speed
Displacement speed along the rail
10 cm/s (4 in/s)
15 cm/s (6 in/s)
20 cm/s (8 in/s)
RTC/RTH
function
RTC/RTH options
Disable
, auto,
,
If RTC/RTH function is set at auto, you need to bring the
unit to the charging station, so the lift will recognize its
location.
For software version 2.0 and 2.1, each time the lift is
turned off, it forgets the charging station location. When
you turn the lift back on, you need to bring it back to the
charging station. For software versions 2.2 and up, even
if the lift is turned off, it will remember the charging sta-
tion’s location.
PPP function
Enabling / Disabling PPP
Disable
, enable
Strap length
RTC/RTH mode strap unwinding
duration
1 to 30 sec. (
1 sec.
) for software version 2.0.X and 2.1.X
13 to 152 cm (5 to 60 in) from software version 2.2.X
Has no effect when RTC/RTH is disabled
Detection time
Maximum time allowed finding Charger
or PPP station
0,5 to 5 min. (
1.5 min.
)
Has no effect when both RTC/RTH and PPP functions
are disabled
Ready state
Lift standby duration before entering
sleep mode
1 to 20 min. (
10.5 min.
)
Hoist led
Enabling / Disabling green LED
Disable,
enable
Date/time
Actual date/time
Depends on value
Set at initial start-up
(
11-01-01 00:00
)
Dual / Single mode
(only for MS2 Plus)
Switching Between Modes
Dual
/ Single
Back
Exit Customizing Mode
Customizing
Back
Monitoring
Monitoring
Back
Customizing
For MS2 Plus the customizing is done on the leading ceiling lift and by doing so the parameters will be set in both the
leading and driven ceiling lifts. For example, if the
Ready
state
setting is changed from 10.5 to 15 minutes, both ceiling
lifts will adopt the changes simultaneously.
Also, parameters
Motor Speed
,
RTC/RTH, PPP
functions and
Detection Time
are found in the menu, customizing them
will have no impact since these functionalities have been disabled within the
Maxi Sky
2 Plus.
001-15697 rev
. 3
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...