8 - 3
HOW TO PERFORM MAINTENANCE
AFTER COMPLETING MAINTENANCE , RESET THE MAINTENANCE CODE;
REFER TO “RESET MAINTENANCE CODE”, PAGE 8-15
NOTE
MAINTENANCE CODE TABLE
PM
Code
Maintenance Description
Trigger
Default
Value
A
Replace the strap
Cycles
5000
B*
Replace the 12V batteries
Cycles
32000
C
Replace real time clock lithium bat-
tery on the main board
Days
1825
D
Annual inspection of the lift
Days
365
E
Annual inspection of spreader bar
Days
365
F
Annual inspection of the rail
installation
Days
365
G
Check KWIKtrak accessory
Malfunction #19 Rail accessory overcurrent
-
H
Check upper limit switch
Malfunction #2 High limit switch fault
-
I
Main board or user interface board
have a memory problem
Malfunction #0: Memory fault
-
J
Replace main board
Malfunction #6 Brake relay fault
Malfunction #12
Main board oscillator fault
Malfunction #13 Unexpected current on Up/Down motor
Malfunction #14 Unexpected current on Left/Right motor
Malfunction #15 Unexpected current on Sit/Reclined motor
Malfunction #17 Accessories or charging defect
Malfunction #20 ECS relay fault
-
L
Encoder Problem
Malfunction #1 Up/Down motor encoder fault
-
M
Problem with Up/Down motor
Malfunction #3 Up/Down motor drive fault
Malfunction #4 Holding system not functional
Malfunction #5 Motor Up/Down wiring fault
Malfunction #7 Overcurrent on Up action
Malfunction #8 Overcurrent on Down action
-
N
Problem with Left/Right motor
Malfunction #9 Overcurrent on Left/Right action
-
O
Problem with Sit/Reclined motor
Malfunction #10 Overcurrent on Sit/Reclined action
-
*
: B PM code is displayed on 2.0 and 2.1 only.
UNLESS OTHERWISE SPECIFIED THE FOLLOWING MAINTENANCE AND INSPECTION INSTRUCTIONS ARE
APPLICABLE FOR BOTH MS2 AND MS2 PLUS.
NOTE
001-15697 rev
. 5
Summary of Contents for Maxi Sky 2
Page 1: ...Maxi Sky 2 Maintenance and Repair Manual with people in mind 001 15697 Rev 5 09 2015 ...
Page 2: ...NOTE This page is left blank intentionally ...
Page 6: ...NOTE This page is left blank intentionally ...
Page 8: ...NOTE This page is left blank intentionally ...
Page 10: ...2 2 NOTE This page is left blank intentionally 001 15697 rev 4 ...
Page 12: ...3 2 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 23: ...4 11 NOTE This page is left blank intentionally 001 15697 rev 5 ...
Page 112: ...7 11 MAXI SKY 2 PLUS C STAT WIRING DIAGRAM 001 15697 rev 5 ...
Page 113: ...7 12 MAXI SKY 2 PLUS ECS WIRING DIAGRAM 001 15697 rev 5 ...
Page 132: ...8 19 TROUBLESHOOTING 001 15697 rev 5 ...
Page 133: ...8 20 001 15697 rev 5 ...
Page 134: ...8 21 001 15697 rev 5 ...
Page 135: ...8 22 001 15697 rev 5 ...
Page 137: ...8 24 001 15697 rev 5 ...
Page 138: ...8 25 001 15697 rev 5 The Troubleshooting section continues on the next page ...